From: Global left ventricular function assessment by ECG-gated multi-detector CT (MDCT): revised role in relation to 2D transthoracic echocardiography

Fig. 4

Bland-Altman plot of LVEF shows the difference between EF by MDCT and 2D echocardiography plotted against the average value of them (solid red line, mean value of difference; green line, mean value of differences ± 2 SDs) a mean value of difference (±SD) of 0.8 ± 11.6% (p < 0.05) between MDCT and 2Dechocardiography. The 95% limits of agreement ranged from −3.3 to 3.2%