Table 4 Comparison of the DRLs of the head and body CT phantom with the corresponding data for real patients study international

From: Image quality and radiation dose assessment for the clinically applied 16-slice CT scanner using PMMA phantom and quality assurance phantom

Study

Year

Country

Design

Dose index

Head (DRLs)

Body (DRLs)

This study

2023

Italy

CT phantom

CTDIvol (mGy)

10

6.1

DLP (mGy*cm)

680

587

Japan

2019

Japan

Human

CTDIvol (mGy)

80

DLP (mGy*cm)

1500

Slovenia

2020

Slovenia

Human

CTDIvol (mGy*cm)

76.9

18.7

DLP (mGy*cm)

1942

1116.2

Egypt

2017

Egypt

Human

CTDIvol (mGy)

30

31

DLP (mGy*cm)

1360

1425

Canada

2016

Canada

Human

CTDIvol (mGy)

82

DLP (mGy*cm)

1302

South India

2018

South India

Human

CTDIvol (mGy)

68

DLP (mGy*cm)

1120

509