Table 3 Quantitative measurement values of the liver in the two groups

From: Evaluation of high temporal resolution magnetic resonance imaging of the liver with gadoxetate disodium in combination with compressed sensing and parallel imaging under single breath-holding using a 1.5-T magnetic resonance system

Phases and measurement items

6-s scan group

10-s scan group

P-value

Multiple arterial

  Mean signal intensity

799.41 ± 166.25 (394.25–1364.62)

767.57 ± 155.50 (481.55–1211.22)

0.275

  SD

62.57 ± 18.91 (35.53–212.43)

60.18 ± 10.57 (39.55–92.40)

0.437

  SNR

13.26 ± 3.05 (4.62–26.93)

12.98 ± 2.79 (7.20–19.16)

0.593

Hepatobiliary

  Mean signal intensity

1232.41 ± 260.40 (633.46–1962.09)

1235.61 ± 248.67 (803.34–1839.16)

0.961

  SD

74.57 ± 16.80 (41.13–129.23)

74.29 ± 15.95 (42.16–110.99)

0.946

  SNR

17.03 ± 4.21 (9.46–32.62)

17.14 ± 3.94 (9.64–25.85)

0.915

  1. Data are presented as mean ± 1 standard deviation. The numbers in parentheses are the ranges
  2. Statistical significance was set at P < 0.05
  3. SD, standard deviation; SNR, signal-to-noise ratio