Skip to main content

Evaluation of different planning methods of 3DCRT, IMRT, and RapidArc for localized prostate cancer patients: planning and dosimetric study

Abstract

Background

Different planning methods of IMRT planning techniques (IMRT (DMLC) and RapidArc) vs. stander techniques (three-dimensional radiation therapy (3DCRT)) will be evaluated for prostate cancer patients’ planning and verification. Three groups of localized prostate cancer patients are planned and evaluated regarding DVHs and practical radiation dosimetry, and ten 3DCRT plans are assessed statistically for each patient.

Results

Plan (7) with parameters of five equally weighted fields with angles of 0°, 45°, 90°, 270°, and 315° and energy of 15MV is the most suitable plan both for PTV coverage and for OAR sparing with a fewer number of fields and fewer number of gantry angles. IMRT complexity involves the requirement of long treatment times and additional effort for planning, safety checks, and quality control before the patient start the treatment and proceed.

Conclusions

The selected plan also is more safe on patients up to 7400 cGy than other plans and is easier to be applied compared to IMRT and RapidArc plans, depending on patient geometry. IMRT radiation doses are more effective and can safely be delivered to PTV with little side effects compared with 3D conformal and conventional techniques. RapidArc has the advantage of re-optimizing and small arcs of variable parameters in dose delivery, taking into account the maximum speed of gantry and MLCs.

Background

Radiation therapy (radical treatment) is widely used for localized prostate curative treatment providing a reduction in late rectal toxicity compared with other conventional treatment techniques [1]. Treatment is carefully planned by using a 3D computed tomography (CT) registered with magnetic resonance (MRI) images of the patient with computerized calculations of doses using treatment planning system (TPS) determining the various intensity patterns and tumor conformal dose [2]. Intensity-modulated radiation therapy (IMRT) is the most suitable treatment modality for the treatment of the lung, prostate, spine, kidney, liver, pancreas, head, and neck cancers. IMRT comes as a solution if it is not available or appropriate for some brain cases (cannot be treated with gamma knife). IMRT is defined as a dose delivery by treatment plans that are optimized using techniques of forward or inverse planning for treatment delivery with modulated beams, using collimators or MLC shaping by modes of either step/shoot (static multileaf collimator (SMLC)) or sliding window (dynamic multileaf collimator (DMLC)) [3]. Techniques of inverse planning included signed compensators for tissue irregularity and missing tissues. Radiation therapy can be capable of destroying all cancer cells, shrinking sizes, or eliminating malignant tumor thus stopping or slowing tumor growth. Radiation is delivered by RapidArcis with gantry rotation as arc continuous on mode; various parameters will be varied (MLC shapes (apertures), dose rate output, gantry speed of rotation, and collimator orientation). RapidArc delivers dose distributions similar to other IMRT techniques or higher [4, 5]. The portal image was obtained from setup fields or treatment beams using a robotic portal imager as a part of radiotherapy treatment machine (linear accelerators (LINACs)) showing that the irradiated area needs to be observed, as it is used for treatment verification, regardless of having lower image quality taken with megavolt irradiation. The portal images’ main use is for the verification of patient set-up, where the patient electronic portal imaging device (EPID) images are matched with anterior and lateral DRR reference images for position verification of patient. Matched images were verified in reference to the patient’s bony landmarks giving the organ position; it is not surely static in the same position relative to the bone. Different ways of verification using information from portal dose are comparing the acquired image and the measured dose image from the portal or by the back projection of information from transmitted dose to calculate patient dose compared to treatment plan dose distribution [6]. These methods may reduce errors in MLC positioning and movement and the LINAC dosimetry and mechanical performance providing a high confidence level for plan accuracy before patient treatment [7].

Objective of the study

This study aims to evaluate different planning methods of IMRT planning techniques (forward planning with three-dimensional radiation therapy (3DCRT), inverse planning with IMRT (DMLC), and RapidArc) for cancer patients. IMRT is very advanced; it needs more quality and accuracy in evaluation before treatment due to many small fields included in plans and different dose areas in the same plan as we have to be sure that the planning encompasses the volumes as prescribed or not. So we did the following:

  1. 1.

    Patient selection and treatment sites (prostate)

  2. 2.

    We studied some images indicative of the tumor

  3. 3.

    Pre-treatment planning

  4. 4.

    Final treatment planning (3DCRT, IMRT, and Rapid Arc) is performed

  5. 5.

    We evaluated different methods of treatment planning:

    • To compare different DVHs for all plans (Eclipse and Prowess planning systems)

    • To use tools of evaluation like biological optimization and physical optimization

    • To conclude and recommend the most suitable treatment methods

Materials and methods

Patients

We choose three groups of prostate patients (20 patients with volumes ranging from 23 to 76 cc). We included a group of patients with intermediate-risk prostate cancer which are patients with clinical stages T2b to T2c, Gleason score 7, or PSA values 10 to 20 ng/ml. CT simulation is performed for all patients with 3-mm slice thickness under the same protocol (positioned in the supine position). Immobilization was done by the knee and ankle support to ensure fixation and reproducibility; the patient arms are on the chests. We scanned the patient by CT from the abdomen till the mid-thigh to allow contouring all the organs at risk. All patients were immobilized with bladder comfortably full and empty rectum; it is a must in every treatment secession as it is done first before CT scan and setup. The clinical target volume (CTV) included the whole prostate gland and the proximal 1 cm of the seminal vesicles. We created PTV by extending the CTV by 1 cm in all directions except posteriorly (only 6 mm) to decrease dose to the rectum, and also, organs at risk (OARs) (bladder, rectum, and both left and right femoral heads) are outlined. The prescribed dose is (74 Gy/7.5 weeks/37 fractions) with a secession dose limit of 200 cGy/Fr.

Beam arrangement

Varian (TrueBeam) linear accelerator with the Eclipse planning system is used. All beam arrangements were visualized using the beam’s eye view (BEV) display. MLC beam shaping was used to create beam apertures. Asymmetric collimation and other parameters were used when necessary with 3DCRT, IMRT, and RapidArc plans.

We were able to deliver dose to the prostate up to 7400 cGy with 3DCRT plans. The X-ray beam energy used was 6 MV or 15 MV when needed regarding the effective path with tissue density correction. Ten 3DCRT plans used as forward plans for each patient with coplanar beam arrangement techniques were designed. Plan (1) was arranged by angles of 0°, 120°, and 240°: anterior direct with an energy of 6 MV and two lateral (Rt and Lt) oblique wedged (W15) beams with an energy of 15MV. Plan (2) was arrange by angles of 0°, 90°, and 270°, anterior direct and two lateral (Rt and Lt) wedged (W60) beams, all beams with an energy of 15MV. Plan (3) was arranged by angles of 0°, 90°, 180°, and 270°: anterior direct with an energy of 6 MV, posterior direct with an energy of 15 MV, and two lateral (Rt and Lt) oblique beams with an energy of 15 MV, all open without wedges. Plan (4) was arranged by angles of 45°, 315°, 135°, and 225°: two anterior oblique wedged beams (W30 L and W30R are directed with the thin end down and the thick end up) with an energy of 6 MV and two posteriors (Rt and Lt) oblique beams with an energy of 15 MV. Plan (5) was arranged by angles of 0°, 45°, 315°, 135°, 180°, and 225°: all fields are open without wedges and with an energy of 15 MV. Plan (6) was arranged by angles of 0°, 60°, 120°, 240°, and 300°: all fields are open without wedges and with an energy of 15 MV. Plan (7) is a five-field technique and was arranged by angles of 0°, 45°, 90°, 270°, and 315°: all fields are open without wedges and with an energy of 15 MV. Plan (8) was arranged by angles of 0°, 60°, 90°, 120°, 220°, 270°, and 300°: one anterior direct beam, two anterior oblique beams with an energy of 6 MV, two direct lateral (Rt and Lt) wedged (W60L and W60R) beams with an energy of 6 MV, and two posterior (Rt and Lt) oblique beams with an energy of 15 MV. Plan (9) is an eight fields technique (with MLC) were arranged by angles of 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°: one anterior direct beam, two anterior oblique beams, two direct lateral (Rt and Lt) wedged (W30 L and W30R are directed with the thin end down and the thick end up) beams, one posterior direct beam, and two posterior (Rt and Lt) oblique, all beams are with an energy of 15 MV. Plan (10) is a plan (9) but without MLC. Plan normalization is set to 100% at beam isocenter, and the PTV is covered by more than 95% isodose; the PTV maximum dose is ranging around 101%. We delivered the dose to the prostate up to 7400 cGy prescription dose by IMRT and RapidArc plans; all beams with an energy of 6 MV. The six-field technique was arranged by angles of 0°, 45°, 90°, 180°, 270°, and 315° with 8 segments per beam (total number of beamlets is 48 openings) and collimator angle of 0°, maximum MU/Fr of 90 MU, and minimum segment area of 1 cm2. The seven-field technique were arranged by angles of 0°, 50°, 90°, 130°, 230°, 270°, and 310° with 6 segments per beam (total number of beamlets is 42 openings) and collimator angle of 0°, maximum MU/Fr of 60 MU, and minimum segment area of 1 cm2. Another seven-field technique was arranged by angles of 0°, 51°, 103°, 155°, 206°, 257°, and 308° with 5 segments per beam (total number of beamlets is 35 openings) and collimator angle of 90°, maximum MU/Fr of 60 MU, and minimum segment area of 1 cm2. While planning, we can increase the penalty weight for both femoral heads until hot spots can reach the acceptable dose limit as hot spots appear near the PTV surface and both femora dose limits may exceed the constraint, taking a long time to achieve optimization. The RapidArc plan was designed with two arcs (179.0° CCW to 181.0° and 181.0° CW to 179.0°) with an energy of 10 MV.

Results

The ten plans were investigated regarding acceptance limits with DVH evaluation, and all results are statistically analyzed before a final decision for an optimal plan that can be accepted with the highest prescribed dose. Figure 1 shows the DVH comparison and plan evaluation for the 3DCRT plan (7) and IMRT plan displaying all organs, showing the plan acceptance, and satisfying all dose limits, saving all OARs; the PTV volume of 95% is covered by 95.9% of dose—saving all of the rectum, bladder, Rt femoral head, and Lt femoral head. The femoral head dose limits used for acceptance are dose maximum (Dmax) < 30Gy, mean dose < 2.5 Gy, and V40 < 40%. The bladder dose limits are Dmax < 65 Gy, V65 ≤ 50%, and V70 ≤ 35%. Rectum dose limits are V50 < 50%, V60 < 35%, V65 < 25%, and V70 < 20%.

Fig. 1
figure 1

Both a 3DCRT plan and b IMRT plan was evaluated according to c dose-volume histogram (DVH) comparison displaying all organs (PTV, rectum, bladder, Rt femoral head, and Lt femoral head)

The RapidArc plan and 3DCRT plan evaluations displaying all organs (PTV, rectum, bladder, Rt femoral head, and Lt femoral head) are shown in Fig. 2. Also, the RapidArc and IMRT plan evaluations with DVH comparison displaying all organs (PTV 74 Gy, rectum, bladder, Rt femoral head, and Lt femoral head) are shown in Fig. 3.

Fig. 2
figure 2

Both a RapidArc plan and b 3DCRT plan were evaluated according to c dose-volume histogram (DVH) comparison displaying all organs (PTV, rectum, bladder, Rt femoral head, and Lt femoral head)

Fig. 3
figure 3

Both a RapidArc plan and b IMRT plan was evaluated according to c dose-volume histogram (DVH) comparison displaying all organs (PTV, rectum, bladder, Rt femoral head, and Lt femoral head)

The IMRT and RapidArc plans achieved the following constraints regarding prostate PTV: penalize all points outside 7644.00 cGy and 8034.00 cGy with the weight of 100.00, penalize if < 99.00% of the volume is below 7644.00 cGy with the weight of 100.00, and penalize if > 1.00% of the volume is above 8190.00 cGy with the weight of 100.00. But for critical organs like bladder, penalize if > 15.00% of volume is above 8002.80 cGy with a weight of 20.00, penalize if > 25.00% of volume is above 7503.60 cGy with a weight of 20.00, penalize if > 50.00% of volume is above 6497.40 cGy with a weight of 20.00, penalize if > 20.00% of volume is above 5499.00 cGy with a weight of 20.00, and penalize if > 1.00% of volume is above 7503.60 cGy with a weight of 78.00. And for rectum, penalize if > 15.00% of the volume is above 7503.60 cGy with the weight of 40.00, penalize if > 25.00% of the volume is above 6497.40 cGy with the weight of 40.00, and penalize if > 50.00% of the volume is above 5499.00 cGy with the weight of 40.00, as rectum constraints are used if the rectum is delineated from the anus to the sigmoid and also if the rectum intersects with PTV with too high volume (< 20% of the rectum volume). For Rt femur, penalize all points > 4500.60 cGy with the weight of 50.00 and penalize all points > 3502.20 cGy with the weight of 20.00. And for Lt femur, penalize all points > 4500.60 cGy with the weight of 50.00 and penalize all points > 3502.20 cGy with the weight of 20.00.

Discussions

Statistical analysis and qualitative analysis for PTV and OAR with delivered dose by the ten 3D plans are discussed. DVH analysis for the ten plans regarding PTV (min, max, mean, V90, and V95 doses) is assessed in Fig. 4 as organs at risk are not exceeding their tolerances [8]. From the statistical analysis, the best dose coverage regarding PTV min is for plan (6) that shows the mean dose of 6648 cGy, but regarding PTV mean dose, we found that the best dose coverage is for plan (9) with the mean of 7437 cGy and the plan (5) mean dose is 99% within the interval between 98% and 99.7%, showing the accepted dose coverage for PTV with fewer number of fields helping in easier and faster dose delivery. Regarding the PTV dose of 95%, the best result is for plan (7) showing the mean of 7178 cGy within the interval of 7098 cGy–7256 cGy with a p value of 0.4, but also, plan (5) shows the mean dose of 7055 cGy (95% of dose) within the interval between 94.2% and 96.5% of the total dose.

Fig. 4
figure 4

DVH analysis for the ten plans regarding PTV (min, max, mean, V90, and V95 doses)

Plan (7) is the best for PTV dose coverage with a 95% isodose line of 7178 ± 117 cGy, 95% CI of 7098–7256, and P = 0.4. Plan (10) is the lowest PTV dose coverage with a 95% isodose line of 7031 ± 257 cGy, 95% CI of 6858–7203, and P = 0.4. Plan (9) is the best for PTV with a mean dose of 7437 ± 125 cGy, 95% CI of 7353–7520, and P = 0.01. Plan (9) is the lowest PTV with a mean dose of 7232 ± 105 cGy, 95% CI of 7152–7292, and P = 0.01. Also, plan (5) is covered by 99% mean dose of 7328 ± 78 cGy, 95% CI of 7275–7380, and P = 0.01; this plan is accepted for treatment, saving much time with a lower number of fields. Regarding PTV minimum dose, plan (6) has the highest PTV dose coverage of 6648 ± 248 cGy, 95% CI of 6481–6815, and P = 0.01. Plan (10) has the lowest PTV minimum dose coverage of 5873 ± 489 cGy, large scale 95% CI of 5545–6201, and P = 0.01. The DVH analysis for the ten plans regarding the bladder (max, mean, V70, V65, V60, and V50 doses) is assessed in Fig. 5a. Plan (7) showed the lowest dose to the bladder with a maximum dose of 6298 ± 2301 cGy, satisfying the condition of maximum dose < 65 Gy, and plan (5) also satisfied the same condition for some plans depending on the prostate volume with P = 0.32. All the ten plans satisfied the critical dose limits of (V70 ≤ 35% and V65 ≤ 50%) as all plans are accepted for treatment. Plan (7) showed acceptable results for both the right and left femoral heads; the Lt femur maximum dose is 3036 ± 731 cGy, with 95% CI of 2545–3527 and P = 0.00, the Lt femur mean dose is 1030 ± 698 cGy, with 95% CI of 560–1499 and P = 0.00, the Rt femur maximum dose was 2814 ± 700 cGy, with 95% CI of 2343–3283 and P = 0.00, and the Rt femur mean dose is 953 ± 456 cGy, with 95% CI of 647–1259 and P = 0.00.

Fig. 5
figure 5

DVH analysis for the ten plans regarding a the bladder (max, mean, V70, V65, V60, and V50 doses) and b rectum (max, mean)

Plan (5) showed acceptable doses for the right and left femoral heads with P = 0.00; the Lt femur maximum dose is 3320 ± 180 cGy with 95% CI of 2108–4530, the Lt femur mean dose is 566 ± 623 cGy with 95% CI of 147–985, and the Rt femur mean dose is 643 ± 377 cGy with 95% CI of 389–896. But plan (7) showed better results than plan (5). All plans satisfied the femoral head dose limit of V40 < 40 Gy. All plans satisfied the tolerance table dose constraints for rectum dose limit of V75 < 15%, and the rectum dose was accepted for treatment as the rectum was empty (volume range is 29.8–100.6 cm2), saving the rectum and reducing rectal toxicity as seen in Fig. 5b. IMRT radiation doses are more effective and can safely be delivered to planning target volumes (PTV) with little side effects compared with 3D conformal and conventional techniques. IMRT complexity involves the requirement of long treatment times and additional effort for planning, safety checks, and quality control before the patient starts the treatment and proceed. Increasing the total radiation dose to the tumor with more sparing for normal tissues and also with more conformity for PTV can be achieved using IMRT better than using conventional or conformal radiotherapy [9]. The overall treatment aims to achieve the goal of full tumor control during every fraction with cancer patients [10, 11]. With the presence of metastatic cancers or continues growth of tumor with other added symptoms, the patient may die due to any failure in tumor full control. All plans were verified using practical radiation dosimetry systems like EPID (for 3DCRT, IMRT) with an acceptance range of 95–99.1% and Delta4 (for RapidArc) with an acceptance range of 98.5–100%) [7, 10, 12].

Conclusion

IMRT radiation doses are more effective and can safely be delivered to planning target volumes (PTV) with little side effects compared with 3D conformal and conventional radiotherapy techniques as the ratio of normal tissue dose to tumor dose is reduced. IMRT complexity involves the requirement of long treatment times and additional effort for planning, safety checks, and quality control before the patient starts the treatment and proceed. Using IMTR helps us to achieve the desire to increase tumor doses and decrease dose to normal organs. The limited use of both IMRT and RapidArc plans is proportional to the high cost of required equipment, operations, and experienced team costs. For prostate treatment, plan (7) with parameters of five equally weighted fields with angles of 0°, 45°, 90°, 270°, and 315° and an energy of 15MV is the most suitable plan both for PTV coverage and for OAR sparing and also saving more time as all beams are open without using wedges as fewer number of fields and fewer number of movements (gantry angles) save a lot of time during treatment, saving patients’ time and saving the treatment center time as well as giving the chance for other patients to be treated. The selected plan also is more safe on patients up to 74 Gy compared to IMRT and RapidArc plans, reducing normal tissue-irradiated volume and reducing toxicity as one of IMRT disadvantages is the presence of secondary malignancies due to the large irradiated volume with low doses during treatment. RapidArc has the advantage of re-optimizing and small arcs of variable parameters in dose delivery, taking into account the maximum speed of gantry and MLCs.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

3DCRT:

Three-dimensional radiation therapy

BEV:

Beam’s eye view

CCW:

Counterclockwise

CT:

Computed tomography

CTV:

Clinical target volume

CW:

Clockwise

D max :

Dose maximum

DMLC:

Dynamic multileaf collimator

DRR:

Digitally reconstructed radiograph

DVH:

Dose-volume histogram

EPID:

Electronic portal imaging device

Fr:

Fraction

Gy:

Gray

IMRT:

Intensity-modulated radiation therapy

LINAC:

Linear accelerators

Lt:

Left

MU:

Monitor unit

OAR:

Organ at risk

PSA:

Prostate-specific antigen

PTV:

Planning target volume

Rt:

Right

SMLC:

Static multileaf collimator

TPS:

Treatment planning system

References

  1. Di Franco R, Borzillo V, Ravo V, Ametrano G, Falivene S, Cammarota F, Rossetti S, Romano FJ, D'Aniello C, Cavaliere C, Iovane G, Piscitelli R, Berretta M, Muto P, Facchini G (2017) Rectal/urinary toxicity after hypofractionated vs conventional radiotherapy in low/intermediate-risk localized prostate cancer: systematic review and meta-analysis. Oncotarget 8(10):17383–17395

    Article  Google Scholar 

  2. Pinnaduwage DS, Descovich M, Lometti MW, Varad B, Roach M, Gottschalk AR (2017) An evaluation of robotic and conventional IMRT for prostate cancer: potential for dose escalation. Technol Cancer Res Treat 16(3):267–275

    Article  Google Scholar 

  3. Salimi M, Abi KST, Nedaie HA, Hassani H, Gharaati H, Samei M, Shahi R, Zarei H (2017) Assessment and comparison of homogeneity and conformity indexes in step-and-shoot and compensator-based intensity modulated radiation therapy (IMRT) and three-dimensional conformal radiation therapy (3D CRT) in prostate cancer. J Med Signals Sensors 7(2):102–107

    Google Scholar 

  4. White P, Chan KC, Cheng KW, Chan KY, Chau MC (2013) Volumetric intensity-modulated arc therapy vs conventional intensity-modulated radiation therapy in nasopharyngeal carcinoma: a dosimetric study. J Radiat Res 54(3):532–545

    Article  Google Scholar 

  5. Huang B, Fang Z, Huang Y, Lin P, Chen Z (2014) A dosimetric analysis of volumetric- modulated arc radiotherapy with jaw width restriction vs 7 fields intensity-modulated radiotherapy for definitive treatment of cervical cancer. Br J Radiol 87(1039):20140183

    Article  CAS  Google Scholar 

  6. Huang M, Huang D, Zhang J, Chen Y, Xu B, Chen L (2017) Preliminary study of clinical application on IMRT three-dimensional dose verification-based EPID system. J Appl Clin Med Phys 18(4):97–105

    Article  Google Scholar 

  7. Kim J, Park S-Y, Kim HJ, Kim JH, Ye S-J, Park JM (2014) The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT. Radiat Oncol (London, England) 9:167

    Article  Google Scholar 

  8. Lu JY, Wu LL, Zhang JY, Zheng J, Cheung ML, Ma CC, Xie LX, Huang BT (2015) Improving target dose coverage and organ-at-risk sparing in intensity-modulated radiotherapy of advanced laryngeal cancer by a simple optimization technique. Br J Radiol 88(1046):20140654

    Article  Google Scholar 

  9. Betzel GT, Yi BY, Niu Y, Yu CX (2012) Is RapidArc more susceptible to delivery uncertainties than dynamic IMRT? Med Phys 39(10):5882–5890

    Article  Google Scholar 

  10. Cozzolino M, Oliviero C, Califano G, Clemente S, Pedicini P, Caivano R, Chiumento C, Fiorentino A, Fusco V (2014) Original paper: clinically relevant quality assurance (QA) for prostate RapidArc plans: gamma maps and DVH-based evaluation. Phys Med 30:462–472

    Article  CAS  Google Scholar 

  11. Vieillevigne L, Molinier J, Brun T, Ferrand R (2015) Gamma index comparison of three VMAT QA systems and evaluation of their sensitivity to delivery errors. Phys Med 31:720–725

    Article  Google Scholar 

  12. Son J, Baek T, Lee B, Shin D, Park SY, Park J, Lim YK, Lee SB, Kim J, Yoon M (2015) A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy. RadiolOncol 49:307–313

    CAS  Google Scholar 

Download references

Acknowledgements

All of the authors thank the anonymous reviewers for their helpful comments on the original manuscript. Also, we thank Dr Ahmed Ramadan Eldesoky ‘MD’ (Clinical Oncology and Nuclear Medicine Department, Faculty of Medicine, Mansoura University) for his help.

Funding

This study had no funding from any resource.

Author information

Authors and Affiliations

Authors

Contributions

KME and ASS carried out the planning processes and studies, participated in the sequence alignment, and drafted the manuscript. MFA selected the patients for the study and carried out the planning target volumes and organ at risk delineation. MFA and ASS performed the statistical analysis. NAB participated in the sequence alignment and participated in the drafting of the manuscript. MTA conceived of the study, participated and supervised in the manuscript design and coordination, and helped to draft it. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Moustafa T. Ahmed.

Ethics declarations

Ethics approval and consent to participate

It is “not applicable” as we did not try the treatment on real patients, and it is only a dosimetric study on a computer.

Consent for publication

It is “not applicable.” We did not ask all the patients included in this research to give written informed consent to publish the data contained within this study as we used their data only for research without application of plans on them (dosimetric study on planning system only).

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Shawata, A.S., Akl, M.F., Elshahat, K.M. et al. Evaluation of different planning methods of 3DCRT, IMRT, and RapidArc for localized prostate cancer patients: planning and dosimetric study. Egypt J Radiol Nucl Med 50, 23 (2019). https://doi.org/10.1186/s43055-019-0021-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s43055-019-0021-z

Keywords