A review on chest CT scanning parameters implemented in COVID-19 patients: bringing low-dose CT protocols into play

Abstract

Background

This study aims to review chest computed tomography (CT) scanning parameters which are utilized to evaluate patients for COVID-19-induced pneumonia. Also, some of radiation dose reduction techniques in CT would be mentioned, because using these techniques or low-dose protocol can decrease the radiation burden on the population.

Main body

Chest CT scan can play a key diagnostic role in COVID-19 patients. Additionally, it can be useful to monitor imaging changes during treatment. However, CT scan overuse during the COVID-19 pandemic raises concerns about radiation-induced adverse effects, both in patients and healthcare workers.

Conclusion

By evaluating the CT scanning parameters used in several studies, one can find the necessity for optimizing these parameters. It has been found that chest CT scan taken using low-dose CT protocol is a reliable diagnostic tool to detect COVID-19 pneumonia in daily practice. Moreover, the low-dose chest CT protocol results in a remarkable reduction (up to 89%) in the radiation dose compared to the standard-dose protocol, not lowering diagnostic accuracy of COVID-19-induced pneumonia in CT images. Therefore, its employment in the era of the COVID-19 pandemic is highly recommended.

Background

New coronavirus (commonly known as COVID-19) pneumonia, which was first reported in Wuhan, Hubei Province, China, in December 2019, followed rapid spread across China and worldwide [1]. This virus is highly contagious and can be spread from person to person by either an infected person or an asymptomatic carrier. The virus can spread in communities and cross borders rapidly. Respiratory droplets (in most of the cases), close contact, and transmission via the digestive tract are possible routes of transmission [2, 3]. Thus far, no curative drug or effective vaccine for COVID-19 is available; therefore, diagnosing the disease at an early stage and to quarantine the infected patients are the most rational approaches to alleviate. China government has published a guideline on Diagnosis and Treatment of Pneumonitis Caused by 2019-nCoV; its latest version says reverse transcription-polymerase chain reaction (RT-PCR) or gene sequencing for samples extracted from the respiratory tract or blood is needed to confirm the diagnosis of COVID-19, which in turn is the main indicator of the need for hospitalization [4]. Though, limitations related to kit performance, sample extraction and sample transportation resulted in about 30 to 60% of RT-PCR tests done on throat swab samples to be positive at initial presentation. This lack of sensitivity will cause many COVID-19 patients not to be identified, leaving a larger population at risk of infection [5].

Imaging also plays a crucial role in the follow-up of novel coronavirus-infected pneumonia (NCIP) patients. Computed tomography (CT) is regarded as the preferable imaging modality in clinically suspected patients and is advantageous to monitor patients during treatment [6, 7]. This imaging modality potentially can spot patients who are highly suspicious of NCIP, regarding clinical conditions with false-negative RT-PCR results [8, 9]. The findings obtained from the CT scan may also provide information on the disease severity [10,11,12]. Moreover, considering similarities in diagnostic patterns of most viral pneumonia, it has been reported in many recent studies that the information derived from CT might be helpful to discriminate COVID-19 from other pathogens causing pneumonia [13]. Chest CT scan (about 56–98% sensitive in COVID-19 detection) is relatively easy to perform and let clinicians diagnose pneumonia fast enough to make important therapeutic decisions effectively early in the disease process [11, 14].

The ionizing radiation exploited in medical imaging is one of the main sources of radiation exposure. It has been reported that the growing popularity and usage of imaging modalities which put ionizing radiation to use leads to an increase in cancer risk [15]. According to the linear no-threshold (LNT) model, there is a direct relationship between the incidence of solid cancers and ionizing radiation, even at low doses. The LNT states that any minimal amount of radiation exposure can increase the incidence of mutation or cancer [16, 17]. The number and dose of CT studies are one of the largest sources of imaging exposure. A typical chest CT scanning delivers a dose of 4–7 mSv to the patient [18], and it raises the risk of cancer incidence and mortality [19, 20]. Thus, although CT scan has played an important role in the management of COVID-19-infected patients, it can promote malignancy. Given that CT scans are mainly performed in patients with clinical suspicion on COVID-19 pneumonia as a standard method to make decisions on hospitalization or discharge, it is possible for each patient suspicious of COVID-19 to undergo one (or more) CT scan(s). Due to the large and uprising number of COVID-19-infected patients worldwide, the risk of developing various types of cancers is worrisome [18]. Considering the growing number of chest CT scans requested by referring physicians and the necessity of decreasing the potential risks posed on patients by ionizing radiation, a dose reduction protocol is an essential requirement [21]. Following the ALARA (as low as reasonably achievable), recommended by the International Commission of Radiological Protection (ICRP), during the daily practice of radiology is necessary, even in the setting of pandemic events [22]. For that reason, we should give thought to optimizing CT protocols implemented in COVID-19-infected patients, to consequently expect a reduction in the patient’s radiation dose.

In this study, we will cover the following sections: (1) chest CT scanning parameters employed so far in patients with COVID-19 infection, (2) CT scanning parameters which can affect the radiation dose (such as scan length, tube current, tube potential, pitch factor, etc.), and (3) low-dose chest CT protocols. We will also review the studies featuring low-dose CT protocols for chest imaging to detect the COVID-19 pneumonia.

Main text

Chest CT scanning parameters used in the diagnosis of COVID-19 infection

Some studies have reported the chest CT findings of the patients infected with COVID-19. In some of these studies, CT scanning parameters have been mentioned, which are extracted and enlisted in Table 1.

Table 1 The data extracted from a number of studies conducted on patients infected with COVID-19: focusing on chest CT scanning parameters

As shown in Table 1, tube voltage is the same in all adult studies (120 kV). Other scanning parameters (such as tube current, pitch factor, and slice thickness) differ from one study to another. In a study by Wang et al. [26], chest CT features of confirmed COVID-19 patients were evaluated, in Xiaogan, Hubei, China. In that study, all patients underwent chest CT examination with a tube current of 320 mA, tube voltage of 120 kV, pitch factor of 1–1.5 (for severe and critical cases), and collimator width of 0.5–1.5 mm. In conclusion, they mentioned that a spiral CT scan, as a sensitive diagnostic tool, can be used to detect early disease and assess its progression. Also, they reported that the spiral chest CT scan is more accurate and sensitive compared to nucleic acid detection in COVID-19 diagnosis [26]. In another study, Song et al. [12] investigated chest CT findings of confirmed COVID-19 patients using RT-PCR in Shanghai, China. CT scanning parameters in their study were as follows: tube voltage of 120 kV, tube current of 180–400 mA, section thickness of 5 mm, and pitch factor of 1.5 [12]. Shen et al. [29] validated the performance of a developed computer tool in order to quantitatively evaluate COVID-19 pneumonia using CT images by comparing the data obtained from computer and radiologist’s notes. In that study, CT scanning parameters were as follows: tube voltage of 120 kV, tube current 30–140 mA, and slice thickness of 0.625 and 1 mm [29]. As mentioned earlier, one of the CT scanning parameters that varies remarkably between different studies is the current tube, which can affect the amount of dose absorbed by the patient (described in detail in the next section).

Moreover, the CT scanning parameters used to diagnose the COVID-19 infection can be different, depending on the studied population. For example, Liu et al. [28] assessed CT features in pregnant women, non-pregnant adults, and children with COVID-19 pneumonia. In that study, the tube voltage used for pregnant women and non-pregnant adults was 120 kV, but it was 100 kV for children. Also, automatic tube current used for pregnant women, non-pregnant adults, and children were 10–300, 120–380, and 30–100 mA, respectively. Helical pitch factor for pregnant women, non-pregnant adults, and children were 1.0875:1 or 1.375:1, 1.375:1, and 0.969:1, respectively [28]. In another study, Xia et al. [25] investigated chest CT findings in pediatric patients with COVID-19 infection. Reported CT scanning parameters in that study were as follows: tube voltage of 12.0 kV, tube current of 100 to 150 mA, collimation of 0.6 mm, and pitch factor of 1:1. As stated in latter two studies [25, 28], the value of scanning parameters used for pregnant women and children were quantitatively lesser than those of adults, an effort to reduce the radiation dose in these age groups. Regarding the radiosensitivity of fetuses and pediatric patients, attention should be drawn to radiation protection principles and dose reduction techniques in CT.

Radiation dose reduction techniques in computed tomography

Before assessment of the scanning parameters affecting the radiation dose in CT, it is necessary to introduce the dose indices in the CT scan.

The primary dose descriptor for CT is computed tomography dose index (CTDI). The CTDI describes the summation of all dose contributions along the z-axis [33, 34]:

$$ {\mathrm{CTDI}}_{100}=\frac{1}{N.T}.{\int}_{-50\ mm}^{+50\ mm}D(z). dz $$
(1)

where D(z) is the value of the dose at a given location, z, and N.T is the nominal beam width [34, 35]. The CTDI100 is measured in the center (CTDI100c) and periphery (CTDI100p) of a polymethyl methacrylate (PMMA) phantom with an ionization chamber with an active length of 100 mm.

Other dose indices for CT dosimetry are weighted CTDI (CTDIw), volumetric CTDI (CTDIv), and dose length product (DLP). The DLP indicates the total amount of radiation received by the patient in a CT examination. The DLP is directly related to the CTDIv (mGy) and length of the scan area (cm) along the longitudinal axis (z-axis) and is measured in milli-gray × cm (mGy × cm). It eases comparison between the amount of radiation exposure in similar examinations [36]. Table 2 presents mathematical expressions of these indices which are well-defined in almost all CT scanners.

Table 2 Common CT dose descriptors

Different organs respond differently to radiation. Given the same amount of radiation, different damages in different organs will ensue. Effective dose measures biologic tissue sensitivity to ionizing radiation effects and is expressed in millisievert. It is the summation of weighted organ dose values (DT,R) and is calculated as follows [39]:

$$ E=\sum {w}_{\mathrm{T}}\sum {w}_{\mathrm{R}}{D}_{\mathrm{T},\mathrm{R}} $$
(2)

where WR is the radiation weighting value and WT is the tissue weighting factor. For diagnostic imaging X-rays, WR is 1. First time, tissue weighting factors were introduced in ICRP report 26. The tissue weighting factors had been updated two times in ICRP 60 and ICRP 103 reports [35].

CT scanning parameters affecting the radiation dose

There are multiple physical and technical dose reduction strategies to reduce radiation dose in CT imaging. Of those, we will review some scanning parameters and equipment-related factors here. Table 3 represents some equipment-related and application-related factors affecting radiation dose in CT (for further details, the reader can refer to Tack et al. [34]).

Table 3 Factors affecting radiation dose in CT [34]

Scan length

Optimization of patient positioning and scan length adjustment avoids radiation to tissues beyond the region of interest and thereby decreases the absorbed dose without any adverse effect on the image quality [40, 41]. As indicated in Table 2, the DLP is directly proportional to scan length. Hence, the patient dose can be decreased by reducing scan length.

Tube current

Adjusting the tube current (mA) is another measure, which results in reduced radiation in the CT scan [42]. The tube current is in proportion to the number of electrons moving from the cathode to anode and thereby is proportional to the quantity of the X-ray emitted from the anode. With tube current increasing, the radiation dose increases linearly. For example, 50% reduction in tube current leads to a 50% reduction in radiation dose [43]. Zarb et al. [44] reported a dose reduction of 20–33%, with a 20–33% reduction in the tube current. Reid et al. [45] measured radiation dose at different tube current-time (mAs) (50–400 mAs) levels in phantom. They also confirmed a linear relationship between tube current and radiation dose.

On the other hand, reducing mAs (proportional to the number of photons absorbed by the detector) leads to decreased image quantum noise (σ) [46]. Image noise is inversely proportional to the square root of mAs:

$$ \sigma =1/\sqrt{\mathrm{mAs}} $$
(3)

It is noteworthy that in order to reduce CT image noise, some methods such as the use of iterative reconstruction algorithms and image filters are proposed which would be discussed later.

Automatic exposure control

Automatic exposure control (AEC) is growing in popularity as an important function in CT scanners, the same exploit as it functions in radiograph machines and has been promising for radiation dose reduction [47]. The main principle of the AEC system is to optimize CT examinations through adapting the tube current to the patient’s size, shape, and attenuation [48]. In CT scanners, the AEC system makes programmed and dynamic tube current adjustment possible, which in turn preserves image quality between patients and for a single patient [49]. The modulation of the tube current can occur in the X-Y plane (angular modulation), along the Z-direction (longitudinal modulation), and in the X-Y-Z plane (combined modulation) [50, 51]. Gudjonsdottir et al. [52] reported that the proper use of AEC can decrease radiation dose values by 15 to 40%. Singh et al. [53] reported 50 to 75% radiation dose reduction could be achieved with the combined modulation technique in chest and abdomen CT in children.

Tube potential

Tube potential is the peak potential (kVp) applied to the X-ray tube, which reflects the energy of the X-ray photons. It has been proven that reduction of the kVp will result in the reduction of the radiation dose. In contrary to the tube current, the tube potential does not linearly correlate with the radiation dose [42, 54]. Huda et al. [54] reported the output values of 1.5 at 100 kVp, 2.5 at 120 kVp, and 3.4 at 140 kVp compared to the 80 kVp. Hamberg et al. [55] measured CTDIw at 80, 100, 120, and 140 kVp and stated that absorbed radiation dose increases as tube voltage to the power of 2.5 ± 0.1 in head mode and 2.8 ± 0.1 in body mode. Zarb et al. [44] examined the effect of reducing the tube potential on radiation dose and image quality parameters for head, abdomen, and chest CT examinations in four CT scanners. They reported that a 14–17% reduction in kVp can result in a 32–38% dose reduction; for instance, by 15% reduction of the kVp (130 to 110 kVp), the CTDI reduces by 38% (7.56 to 4.69 mGy). Kubo et al. [40] reviewed dose reduction strategies in chest CT and concluded that using lower kVp is particularly useful for pediatric patients and small size adults weighing less than 55 kg. They suggested taking advantage from the Care KV, provided by the Siemens for the tube voltage automatic selection, to overcome this problem.

Pitch factor

Pitch factor (simply pitch) is defined as the ratio of the CT table increment distance during one gantry rotation to the total beam width. For single-slice CT scanners, increasing the pitch leads to a decrease in the radiation dose proportionally. Mahesh et al. [56] reported radiation dose values of 12.72, 6.68, and 3.62 mGy for pitch factors of 0.5, 1, and 2, respectively. In contrast, for multi-slice helical CT scanners, radiation dose is independent of pitch factor. They reported radiation dose values of 9.92, 9.94, and 10.12 mGy for pitch factors of 2, 4, and 8, respectively. An explanation for this is the conception of effective mAs, so that the effective mAs or mAs per slice is the actual mAs value divided by the pitch (effective mAs = mAs/pitch) [57]. As the pitch increases in multi-slice CT scanners, the noise will increase with pitch and, subsequently, the CT scanner automatically increases mAs (to keep constant effective mAs) to prevent image quality degradation. This type of adjustment is available almost in all multi-slice CT scanners [34]. If one applies constant mAs, the radiation dose will decrease with increasing the pitch factor. Therefore, the CT users must gain enough knowledge on CT scanners they are working with and beware of the differences between mA and effective mA.

Reconstruction algorithm and image filter

Dose reduction in CT can be achieved by reducing the tube current (mA) or tube potential (kVp) or combination of both. Both of these methods will increase image noise. Iterative image reconstruction methods have been introduced in order to decrease the image noise in CT images. Yanagawa et al. [58] investigated the effects of adaptive statistical iterative reconstruction (ASIR) method on the computer-aided detection (CAD) system in pulmonary nodules using routine CT protocol and low-dose CT. They investigated 34 patients on 64-slice multi-detector CT scanner with 120 kVp at 100 mAs (routine CT protocol) and automatically adjusted mA (lower dose CT) with three different combinations of filtered back-projection and ASIR algorithms: 0% (pure filtered back-projection), 50%, and 100% (pure ASIR). The images were reviewed by three radiologists, and their results showed a significantly higher CAD sensitivity at 100% ASIR algorithm compared to 0% ASIR algorithm. Also, the effective dose was significantly higher in routine CT protocol compared to lower dose CT (10.77 mSv vs. 2.67 mSv) [58].

Another way of image noise reduction is using image filters (or kernels). Image filters can be applied on raw data or the reconstructed images. In a phantom study, Khoramian et al. [59] investigated the effect of different reconstruction kernels on noise and spatial resolution of CT images. In that study, an image was obtained from a conventional quality control phantom and it was reconstructed with five different kernels on Siemens Emotion 6 CT scanner. The employed kernels were H10, H30, H40, H50, and H70 (“H” indicates the scanned region, i.e., the head, and the numbers represent the softness and sharpness of the kernels, as H10 and H70 are the softest and the sharpest filters, respectively). Their findings indicated that the image noise and the spatial resolution are influenced by the kernel, as from H10 to H70, the image noise increases from 2.61 to 19.59, while the spatial resolution will get improved from 1.80 to 1.04 mm [59].

Low-dose CT protocol

There is no quantitative well-defined low-dose CT protocol. However, the aim of this protocol is to reduce the radiation dose delivered to the patient by changing radiation parameters mainly kVp and mAs. In fact, the main goal of low-dose CT protocol is to achieve acceptable diagnostic accuracy while keeping radiation dose at lower levels. One of the first efforts in performing low-dose chest CT was done by Naidich et al. [60] in 1990. They investigated and compared chest CT images of 12 patients between standard protocol and low-dose CT. Among study cases, two were normal and ten cases had evidence of diffuse lung disease. Two CT protocols were used and the images were evaluated independently by two experienced chest radiologists. The first set of CT scan images was obtained with 140 mA, 120 kV, 2-s scan time, 10 mm thickness, and standard construction algorithm. Low-dose protocol differed from the standard protocol only in tube current (10 mA). In conclusion, they reported that visualization of parenchymal structures was not affected by reducing mA [60].

Rusinek et al. [61] assessed the effectiveness of low-dose CT in the identification of pulmonary nodules. Six radiologists with 1 to 16 years of experience reviewed 864 (432 taken with the low-dose protocol and 432 taken with the conventional protocol) positive images and rated a 4-point scale for each image. Conventional scans were obtained at 120 kVp and 200 mAs with a slice thickness of 10 mm. The low-dose images were acquired with 20 mAs but all other scan parameters were the same as conventional images. Sensitivities of conventional and low-dose CT images were 63% and 60%, respectively. Also, specificities of conventional and low-dose CT images were 91% and 88%, respectively. There were no significant differences between sensitivity and specificity of low-dose and conventional images. Finally, they conclude that the low-dose CT protocol is acceptable for pulmonary nodule identification [61]. Rogalla et al. [62] examined chest low-dose CT examination in pediatric patients done for different indications. Tube voltage of 120 kV and mAs values of 12.5 to 75 per slice were applied. In that study, the diagnostic value of low-dose CT examinations was compared to standard-dose chest CT examinations of adults (tube voltage of 120 kV and 175 mAs per slice) in terms of technical image quality (noise and artifacts). Their results showed that for all indications in the chest CT scanning of children, the low-dose protocol provides sufficient image quality, without loss of diagnostic data. Also, the radiation dose obtained from the low-dose CT protocol was approximately 5–20% of a standard-dose CT protocol [62]. Ono et al. [63] compare image quality and radiation dose between low-dose CT and follow-up standard CT protocols for lung cancer screening in 19 adult patients. Low-dose CT images were obtained with 120 kVp, 30–50 mA, total beam collimation of 16 mm, and pitch of 1.438. Follow-up CT images were taken with 120 kVp, AEC (max 150 mAs), total beam collimation of 16 mm, and pitch of 1.188. There were no significant differences between these two protocols (low-dose CT and follow-up CT) for diagnosing lung cancer. Moreover, effective dose values ranged between 1.3 and 3.4 mSv and 8.5 and 14.0 mSv for low-dose and follow-up CT protocols, respectively. In conclusion, they suggested that low-dose CT can be used as an alternative to follow-up standard CT, an attempt to reduce radiation dose [63]. Kubo et al. [64] compared CT images of 59 patients with two different CT protocols: low-dose CT with 50 mAs at 120 kVp and 150 mAs at 120 kVp (other CT scanning parameters were identical). They found that CTDIw in low-dose CT is significantly lower than that of standard CT (5.36 mGy against 16.1 mGy). On the other hand, there was no significant difference in the deviation of the readers between low-dose CT and standard-dose CT [64].

Perspective of future research

The first patient infected with a novel coronavirus was reported in 2019 December, Wuhan, China, and in less than 6 months (as announced by the World Health Organization), the total number of COVID-19 cases worldwide has crossed 5 million (https://www.who.int/emergencies/diseases/novel-coronavirus-2019/situation-reports).

CT scan role has been approved as a diagnostic tool to help in evaluation, diagnosis, and treatment management in COVID-19-infected patients. Despite the mentioned advantages of chest CT scan, and its uses in making decisions for infected patients hospitalization or discharge, CT scan overuse raises concerns on increased radiation.

According to the LNT model, there is no safe level of radiation exposure and any radiation dose greater than 0 can lead to cancer or genetic mutations [16]. Hence, in X-ray-based medical imaging modalities, reduction in patient’s radiation, with or without negligible change in the image quality, is important [65]. In this regard, following as low as reasonably achievable (ALARA) principles can play a key role in radiation protection [46].

Based on the studies mentioned above, the scanning parameters (such as tube current and tube voltage) which are used to diagnose the COVID-19 pneumonia on CT scan can be optimized pursuing low-dose CT protocols. However, dosimetric indicators (such as CTDI, DLP, effective dose, etc.) were not mentioned in most of the studies and we could not discuss these quantities further. In a recent study by Dangis et al. [66], the accuracy and reproducibility of the low-dose chest CT protocol to diagnose COVID-19 pneumonia were evaluated. Mentioned CT scanning parameters were as follows: tube voltage of 100 kVp, tube current-time of 20 mAs, pitch factor of 1.2, and 0.5 second gantry rotation time. The mean effective radiation dose obtained from the low-dose protocol was submillisievert (0.56 ± 0.25 mSv). Finally, they concluded that low-dose submillisievert chest CT scan can be applied to assess COVID-19 pneumonia rapidly, accurately, and reproducibly [66]. Additionally, Radpour et al. [21] recommended CT scanning parameters of kVp 100–120, mAs 50–100, pitch factor 0.8–1.5, and thickness 1–3 mm, in order to minimize the radiation dose of patients. In conclusion, they stated that a low-dose high-resolution CT protocol might be more advantageous than RT-PCR, particularly in highly infected societies with low availability of PCR kits [21]. Kang et al. [67] also recommended the low-dose CT protocol in the diagnosis and management of COVID-19-induced pneumonia. They evaluated COVID-19 pneumonia in two adults of similar age, height, and body size by standard-dose and low-dose protocols and reported that the image quality in both protocols was excellent; nevertheless, effective dose and DLP values in the low-dose protocol were 0.203 mSv and 14.5 mGy cm, respectively, compared with 1.8074 mSv and 129.1 mGy cm in the standard-dose protocol [67]. In another report, Tofighi et al. [68] represented that chest CT low-dose protocol compared to conventional CT chest protocol reduced radiation dose to less than 50% without a remarkable impact on the diagnostic value. Shiri et al. [69] proposed a low-dose protocol based on deep learning algorithms and reported that using this protocol (kVp 90, mAs 20–40, pitch factor 0.75) compared to full-dose CT protocol (kVp 100–120, mAs 100–150, pitch factor 1.3–1.8) reduced CTDIvol, as a good dosemetric for comparing the effective dose and radiation-induced risks, by up to 89% which reflects a considerable decrease of the radiation dose value associated with CT examinations [69].

From a clinical point of view, recent findings on evaluation of ground-glass and consolidative opacities (as the main criteria in diagnosing COVID-19 pneumonia) showed comparable results for both low-dose and standard-dose CT chest protocols [68]. Also, low-dose chest CT scan images in 5 different patients with RT-PCR-confirmed COVID-19 pneumonia (from our institution [Shahid Beheshti Hospital, Kashan, Iran]; CT scanning parameters: kVp 120, mAs 50, pitch factor 1, thickness 3 mm) show that variety of diseased lung extension and density are comparable with the standard-dose CT scan (Fig. 1).

Fig. 1
figure 1

Low-dose chest CT scan images in 5 different patients with RT-PCR-confirmed COVID-19 pneumonia, showing a variety of diseased lung extension and density, comparable with the standard-dose CT scan. a Peripheral area of ground-glass opacity (thick short arrow) with overlying pleural thickening (thin long arrow). b Mixed central-peripheral lesion of ground-glass opacity (thick short arrow) with a prominent intralesional vessel (thin long arrow). c Peripheral area of ground-glass opacity (black arrows) with prominent intralesional vessels. d Peripheral area of crazy paving pattern (black arrows) with thickened interlobular septa within (blue arrow). e Mixed ground-glass opacity (arrowhead) and consolidation (thick black arrow). Long thin arrows point out air-bronchogram [images are from first author’s institution, implementing low-dose CT scan protocol, with scanning parameters as follows: kVp 120, mAs 50, pitch factor 1, thickness 3 mm]

Also, the results obtained from the low-dose chest CT protocol in the diagnosis of COVID-19 pneumonia [68, 69] were in agreement with low-dose chest CT protocols used in detecting other diseases (for example, in the identification of pulmonary nodules, in lung cancer screening, etc.) [61, 63], as the findings showed that the radiological features resulted from low-dose protocols can be comparable to those obtained from the standard-dose protocols, with a significant reduction in radiation dose.

Conclusion

Given the growing number of chest CT scan done during the COVID-19 pandemic, the increased radiation burden on patients and medical staff is becoming a matter of concern. By evaluating the CT scanning parameters used in several studies, one can find the necessity for optimizing these parameters. The findings obtained from the low-dose chest CT protocol (compared to the standard-dose protocol) showed that using this protocol is reliable in the detection of COVID-19 pneumonia in daily practice. Moreover, the low-dose chest CT protocol results in a remarkable reduction (up to 89%) in the radiation dose compared to the standard-dose protocol, not lowering the diagnostic accuracy of COVID-19-induced pneumonia in CT images. Therefore, its employment in the era of the COVID-19 pandemic is highly recommended.

One of the main limitations of using the low-dose chest CT protocols to detect COVID-19 pneumonia is that there are few studies available which have compared the radiological findings in both low-dose and standard-dose protocols; hence, further comparative studies are suggested to get assured of low-dose CT scan sufficiency in diagnosing COVID-19 pneumonia.

Availability of data and materials

Not applicable.

Abbreviations

ALARA:

As low as reasonably achievable

CT:

Computed tomography

COVID-19:

Coronavirus 2019

2019-nCoV:

2019 novel coronavirus

CTDI:

Computed tomography dose index

DLP:

Dose length product

RT-PCR:

Reverse transcription-polymerase chain reaction

NCIP:

Novel coronavirus-infected pneumonia

LNT:

Linear no-threshold

References

  1. Chen H, Guo J, Wang C, Luo F, Yu X, Zhang W et al (2020) Clinical characteristics and intrauterine vertical transmission potential of COVID-19 infection in nine pregnant women: a retrospective review of medical records. Lancet 395:809–815. https://doi.org/10.1016/S0140-6736(20)30360-3

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  2. Riou J, Althaus CL (2020) Pattern of early human-to-human transmission of Wuhan 2019 novel coronavirus (2019-nCoV), December 2019 to January 2020. Euro Surveill 25:2000058. https://doi.org/10.2807/1560-7917.ES.2020.25.4.2000058

    Article  PubMed Central  Google Scholar 

  3. Tipaldi MA, Lucertini E, Orgera G, Zolovkins A, Lauirno F, Ronconi E et al (2020) How to manage the COVID-19 diffusion in the angiography suite: experiences and results of an Italian interventional radiology unit. Sci Med J 2:1–8. https://doi.org/10.28991/SciMedJ-2020-02-SI-1

    Article  Google Scholar 

  4. General Office of National Health Committee Notice on the issuance of a program for the diagnosis and treatment of novel coronavirus (2019-nCoV) infected pneumonia (trial sixth edition). http://www.nhc.gov.cn/yzygj/pqt/new_list.shtml. Accessed 24 Feb 2020.

  5. Ai T, Yang Z, Hou H, Zhan C, Chen C, Lv W et al (2020) Correlation of chest CT and RT-PCR testing in coronavirus disease 2019 (COVID-19) in China: a report of 1014 cases. Radiology 296:32–40. https://doi.org/10.1148/radiol.2020200642

    Article  CAS  Google Scholar 

  6. Zhao W, Zhong Z, Xie X, Yu Q, Liu J (2020) Relation between chest CT findings and clinical conditions of coronavirus disease (COVID-19) pneumonia: a multicenter study. AJR Am J Roentgenol 214:1072–1077. https://doi.org/10.2214/AJR.20.22976

    Article  PubMed  Google Scholar 

  7. Hanscom D, Clawson DR, Porges SW, Bunnage R, Aria L, Lederman S et al (2020) Polyvagal and global cytokine theory of safety and threat Covid-19–plan B. Sci Med J 2:9–27. https://doi.org/10.28991/SciMedJ-2020-02-SI-2

    Article  Google Scholar 

  8. Xie X, Zhong Z, Zhao W, Zheng C, Wang F, Liu JJR (2020) Chest CT for typical 2019-nCoV pneumonia: relationship to negative RT-PCR testing. Radiology 296:41–45. https://doi.org/10.1148/radiol.2020200343.

    Article  Google Scholar 

  9. Wu J, Wu X, Zeng W, Guo D, Fang Z, Chen L et al (2020) Chest CT findings in patients with coronavirus disease 2019 and its relationship with clinical features. Invest Radiol 55:257–261. https://doi.org/10.1097/RLI.0000000000000670

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  10. Huang C, Wang Y, Li X, Ren L, Zhao J, Hu Y et al (2020) Clinical features of patients infected with 2019 novel coronavirus in Wuhan, China. Lancet 395:497–506. https://doi.org/10.1016/S0140-6736(20)30183-5

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  11. Kanne JP, Little BP, Chung JH, Elicker BM, Ketai LH (2020) Essentials for radiologists on COVID-19: an update—radiology scientific expert panel. Radiology 296:113–114. https://doi.org/10.1148/radiol.2020200527

    Article  Google Scholar 

  12. Song F, Shi N, Shan F, Zhang Z, Shen J, Lu H et al (2020) Emerging 2019 novel coronavirus (2019-nCoV) pneumonia. Radiology 295:210–217. https://doi.org/10.1148/radiol.2020200274

    Article  PubMed  Google Scholar 

  13. Azadbakht J, Haghi-Aminjan H, Farhood B (2020) Chest CT findings of COVID-19-infected patients, are there differences between pediatric and adult patients? A systematic review. Egypt J Radiol Nucl Med 51:1–10. https://doi.org/10.1186/s43055-020-00261-8

    Article  Google Scholar 

  14. Fang Y, Zhang H, Xie J, Lin M, Ying L, Pang P et al (2020) Sensitivity of chest CT for COVID-19: comparison to RT-PCR. Radiology 296:115–117. https://doi.org/10.1148/radiol.2020200432

    Article  Google Scholar 

  15. Shah D, Sachs R, Wilson D (2012) Radiation-induced cancer: a modern view. Br J Radiol 85:1166–1173. https://doi.org/10.1259/bjr/25026140

    Article  Google Scholar 

  16. Mullenders L, Atkinson M, Paretzke H, Sabatier L, Bouffler S (2009) Assessing cancer risks of low-dose radiation. Nat Rev Cancer 9(8):596–604. https://doi.org/10.1038/nrc2677

    Article  CAS  PubMed  Google Scholar 

  17. Aliasgharzadeh A, Mohseni M, Salimian M, Farhood B, Najafi M, Safari H et al (2020) Cumulative effective dose caused by diagnostic imaging and its associated risk for cancer development in trauma patients referred to the emergency department. J Med Sci 40:51–58. https://doi.org/10.4103/jmedsci.jmedsci_14_19

    Article  Google Scholar 

  18. Wang YXJ, Liu W-H, Yang M, Chen W (2020) The role of CT for Covid-19 patient’s management remains poorly defined. Ann Transl Med 8:145. https://doi.org/10.21037/atm.2020.02.71

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  19. Davoudi M, Khoramian D, Abedi-Firouzjah R, Ataei G (2019) Strategy of computed tomography image optimisation in cervical vertebrae and neck soft tissue in emergency patients. Radiat Prot Dosimetry 187:98–102. https://doi.org/10.1093/rpd/ncz145

    Article  PubMed  Google Scholar 

  20. Osipov M, Vazhenin A, Kuznetsova A, Aksenova I, Vazhenina D, Sokolnikov M (2020) PET-CT and occupational exposure in oncological patients. Sci Med J 2:63–69. https://doi.org/10.28991/SciMedJ-2020-0202-3

    Article  Google Scholar 

  21. Radpour A, Bahrami-Motlagh H, Taaghi MT, Sedaghat A, Karimi MA, Hekmatnia A et al (2020) COVID-19 evaluation by low-dose high resolution CT scans protocol. Acad Radiol 27:901. https://doi.org/10.1016/j.acra.2020.04.016

    Article  PubMed  PubMed Central  Google Scholar 

  22. Yeung W (2019) The ‘as low as reasonably achievable’ (ALARA) principle: a brief historical overview and a bibliometric analysis of the most cited publications. Radioprotection 54:111–112. https://doi.org/10.1051/radiopro/2019017

    Article  CAS  Google Scholar 

  23. Yoon SH, Lee KH, Kim JY, Lee YK, Ko H, Kim KH et al (2020) Chest radiographic and CT findings of the 2019 novel coronavirus disease (COVID-19): analysis of nine patients treated in Korea. Korean J Radiol 21:494–500. https://doi.org/10.3348/kjr.2020.0132

    Article  PubMed  PubMed Central  Google Scholar 

  24. Xu X, Yu C, Qu J, Zhang L, Jiang S, Huang D et al (2020) Imaging and clinical features of patients with 2019 novel coronavirus SARS-CoV-2. Eur J Nucl Med Mol Imaging 47:1275–1280. https://doi.org/10.1007/s00259-020-04735-9

    Article  CAS  PubMed  Google Scholar 

  25. Xia W, Shao J, Guo Y, Peng X, Li Z, Hu D (2020) Clinical and CT features in pediatric patients with COVID-19 infection: different points from adults. Pediatr Pulmonol 55:1169–1174. https://doi.org/10.1002/ppul.24718

    Article  PubMed  PubMed Central  Google Scholar 

  26. Wang K, Kang S, Tian R, Zhang X, YJCr W (2020) Imaging manifestations and diagnostic value of chest CT of coronavirus disease 2019 (COVID-19) in the Xiaogan area. Clin Radiol 75:341–347. https://doi.org/10.1016/j.crad.2020.03.004

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  27. Li M, Lei P, Zeng B, Li Z, Yu P, Fan B et al (2020) (2020) coronavirus disease (COVID-19): spectrum of CT findings and temporal progression of the disease. Acad Radiol 27:603–608. https://doi.org/10.1016/j.acra.2020.03.003.

    Article  PubMed  PubMed Central  Google Scholar 

  28. Liu H, Liu F, Li J, Zhang T, Wang D, Lan W (2020) Clinical and CT imaging features of the COVID-19 pneumonia: focus on pregnant women and children. J Inf Secur 80:7–13. https://doi.org/10.1016/j.jinf.2020.03.007

    Article  CAS  Google Scholar 

  29. Shen C, Yu N, Cai S, Zhou J, Sheng J, Liu K et al (2020) Quantitative computed tomography analysis for stratifying the severity of coronavirus disease 2019. J Pharm Anal 10:123–129. https://doi.org/10.1016/j.jpha.2020.03.004

    Article  PubMed  PubMed Central  Google Scholar 

  30. Xu Y-H, Dong J-H, An W-M, Lv X-Y, Yin X-P, Zhang J-Z et al (2020) Clinical and computed tomographic imaging features of novel coronavirus pneumonia caused by SARS-CoV-2. J Inf Secur 80:394–400. https://doi.org/10.1016/j.jinf.2020.02.017

    Article  CAS  Google Scholar 

  31. Albarello F, Pianura E, Di Stefano F, Cristofaro M, Petrone A, Marchioni L et al (2020) 2019-novel coronavirus severe adult respiratory distress syndrome in two cases in Italy: an uncommon radiological presentation. Int J Infect Dis 93:192–197. https://doi.org/10.1016/j.ijid.2020.02.043

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  32. Pan F, Ye T, Sun P, Gui S, Liang B, Li L et al (2020) Time course of lung changes on chest CT during recovery from 2019 novel coronavirus (COVID-19) pneumonia. Radiology 295:715–721. https://doi.org/10.1148/radiol.2020200370.

    Article  PubMed  Google Scholar 

  33. Khoramian D, Sistani S, Hejazi P (2019) Establishment of diagnostic reference levels arising from common CT examinations in Semnan County, Iran. Polish J Med Phys Eng 25:51–55. https://doi.org/10.2478/pjmpe-2019-0008

    Article  Google Scholar 

  34. Tack D, Gevenois PA, Abada H (2007) Radiation dose from adult and pediatric multidetector computed tomography. Springer. https://doi.org/10.1007/978-3-540-68575-3

  35. ICRP Publication 103 (2007) The 2007 recommendations of the international commission on radiological protection: ICRP publication 103. Ann ICRP 37:1–332. https://doi.org/10.1016/j.icrp.2007.10.003

    Article  Google Scholar 

  36. Huda W, Mettler FA (2011) Volume CT dose index and dose-length product displayed during CT: what good are they? Radiology 258:236–242. https://doi.org/10.1148/radiol.10100297

    Article  PubMed  Google Scholar 

  37. Khoramian D, Hashemi B (2017) Effective and organ doses from common CT examinations in one general hospital in Tehran, Iran. Pol J Med Phys Eng 23:73–79. https://doi.org/10.1515/pjmpe-2017-0013

    Article  Google Scholar 

  38. Khoramian D, Sistani S (2017) Estimation and comparison of the radiation effective dose during coronary computed tomography angiography examinations on single-source 64-MDCT and dual-source 128-MDCT. J Radiol Prot 37:826–836. https://doi.org/10.1088/1361-6498/aa823f

    Article  PubMed  Google Scholar 

  39. Brady SL, Mirro AE, Moore BM, Kaufman RA (2015) How to appropriately calculate effective dose for CT using either size-specific dose estimates or dose-length product. AJR Am J Roentgenol 204:953–958. https://doi.org/10.2214/AJR.14.13317

    Article  PubMed  Google Scholar 

  40. Kubo T, Ohno Y, Kauczor HU, Hatabu H (2014) Radiation dose reduction in chest CT—review of available options. Eur J Radiol 83:1953–1961. https://doi.org/10.1016/j.ejrad.2014.06.033

    Article  PubMed  Google Scholar 

  41. Yu L, Liu X, Leng S, Kofler JM, Ramirez-Giraldo JC, Qu M et al (2009) Radiation dose reduction in computed tomography: techniques and future perspective. Imaging Med 1:65–84. https://doi.org/10.2217/iim.09.5

    Article  PubMed  PubMed Central  Google Scholar 

  42. Singh S, Kalra MK, Thrall JH, Mahesh M (2011) CT radiation dose reduction by modifying primary factors. J Am Coll Radiol 8:369–372. https://doi.org/10.1016/j.jacr.2011.02.001

    Article  PubMed  Google Scholar 

  43. Lira D, Padole A, Kalra MK, Singh S (2015) Tube potential and CT radiation dose optimization. AJR Am J Roentgenol 204:4–10. https://doi.org/10.2214/AJR.14.13281

    Article  Google Scholar 

  44. Zarb F, Rainford L, McEntee MF (2011) Developing optimized CT scan protocols: phantom measurements of image quality. Radiography 17:109–114. https://doi.org/10.1016/j.radi.2010.10.004

    Article  Google Scholar 

  45. Reid J, Gamberoni J, Dong F, Davros W (2010) Optimization of kVp and mAs for pediatric low-dose simulated abdominal CT: is it best to base parameter selection on object circumference? AJR Am J Roentgenol 195:1015–1020. https://doi.org/10.2214/AJR.09.3862

    Article  PubMed  Google Scholar 

  46. Khoramian D, Sistani S, Firouzjah RA (2019) Assessment and comparison of radiation dose and image quality in multi-detector CT scanners in non-contrast head and neck examinations. Pol J Radiol 84:61–67. https://doi.org/10.5114/pjr.2019.82743

    Article  Google Scholar 

  47. Matsumoto K, Ohno Y, Koyama H, Kono A, Inokawa H, Onishi Y, Nogami M, Takenaka D, Araki T, Sugimura K (2011) 3D automatic exposure control for 64-detector row CT: radiation dose reduction in chest phantom study. Eur J Radiol 77:522–527. https://doi.org/10.1016/j.ejrad.2009.09.012

    Article  PubMed  Google Scholar 

  48. Söderberg M, Gunnarsson M (2010) Automatic exposure control in computed tomography–an evaluation of systems from different manufacturers. Acta Radiol 51:625–634. https://doi.org/10.3109/02841851003698206

    Article  PubMed  Google Scholar 

  49. Koyama H, Ohno Y, Yamazaki Y, Matsumoto K, Onishi Y, Takenaka D et al (2012) Reduced-dose chest CT with 3D automatic exposure control vs. standard chest CT: quantitative assessment of emphysematous changes in smokers’ lung parenchyma. Eur J Radiol 81:1330–1334. https://doi.org/10.1016/j.ejrad.2011.03.037

    Article  PubMed  Google Scholar 

  50. Söderberg M (2008) Automatic exposure control in CT an investigation between different manufacturers considering radiation dose and image quality DOI: http://lup.lub.lu.se/student-papers/record/2157064

    Google Scholar 

  51. Matsubara K, Koshida K, Noto K, Takata T (2011) Reduction of breast dose in abdominal CT examinations: effectiveness of automatic exposure control system. Radiat Meas 46:2056–2059. https://doi.org/10.1016/j.radmeas.2011.07.018

    Article  CAS  Google Scholar 

  52. Gudjonsdottir J, Ween B, Olsen DR (2010) Optimal use of AEC in CT: a literature review. Radiol Technol 81:309–317 PMID: 21048061

    PubMed  Google Scholar 

  53. Singh S, Kalra MK, Thrall JH, Mahesh M (2011) Automatic exposure control in CT: applications and limitations. J Am Coll Radiol 8:446–449. https://doi.org/10.1016/j.jacr.2011.03.001

    Article  PubMed  Google Scholar 

  54. Huda W, Scalzetti EM, Levin G (2000) Technique factors and image quality as functions of patient weight at abdominal CT. Radiology 217:430–435. https://doi.org/10.1148/radiology.217.2.r00nv35430

    Article  CAS  PubMed  Google Scholar 

  55. Hamberg LM, Rhea JT, Hunter GJ, Thrall JH (2003) Multi–detector row CT: radiation dose characteristics. Radiology 226:762–772. https://doi.org/10.1148/radiol.2263020205

    Article  PubMed  Google Scholar 

  56. Mahesh M, Scatarige JC, Cooper J, Fishman EK (2001) Dose and pitch relationship for a particular multislice CT scanner. AJR Am J Roentgenol 177:1273–1275. https://doi.org/10.2214/ajr.177.6.1771273

    Article  CAS  PubMed  Google Scholar 

  57. McKenney SE, Seibert JA, Lamba R, Boone JM (2014) Methods for CT automatic exposure control protocol translation between scanner platforms. J Am Coll Radiol 11:285–291. https://doi.org/10.1016/j.jacr.2013.10.014

    Article  PubMed  PubMed Central  Google Scholar 

  58. Yanagawa M, Honda O, Kikuyama A, Gyobu T, Sumikawa H, Koyama M, Tomiyama N (2012) Pulmonary nodules: effect of adaptive statistical iterative reconstruction (ASIR) technique on performance of a computer-aided detection (CAD) system—comparison of performance between different-dose CT scans. Eur J Radiol 81:2877–2886. https://doi.org/10.1016/j.ejrad.2011.09.011

    Article  PubMed  Google Scholar 

  59. Khorramian D, Hashemi Malayeri B, Sistani S (2016) Effects of different reconstruction kernels on noise and spatial resolution of computed tomography images: a phantom study. Paramed Sci Military Health 11:25–29 DOI: http://jps.ajaums.ac.ir/article-1-70-en.html

    Google Scholar 

  60. Naidich DP, Marshall CH, Gribbin C, Arams RS, McCauley DI (1990) Low-dose CT of the lungs: preliminary observations. Radiology 175:729–731. https://doi.org/10.1148/radiology.175.3.2343122

    Article  CAS  PubMed  Google Scholar 

  61. Rusinek H, Naidich DP, McGuinness G, Leitman BS, McCauley DI, Krinsky GA et al (1998) Pulmonary nodule detection: low-dose versus conventional CT. Radiology 209:243–249. https://doi.org/10.1148/radiology.209.1.9769838

    Article  CAS  PubMed  Google Scholar 

  62. Rogalla P, Stöver B, Scheer I, Juran R, Gaedicke G, Hamm B (1999) Low-dose spiral CT: applicability to paediatric chest imaging. Pediatr Radiol 29:565–569. https://doi.org/10.1007/s002470050648

    Article  CAS  PubMed  Google Scholar 

  63. Ono K, Hiraoka T, Ono A, Komatsu E, Shigenaga T, Takaki H et al (2013) Low-dose CT scan screening for lung cancer: comparison of images and radiation doses between low-dose CT and follow-up standard diagnostic CT. SpringerPlus 2:393. https://doi.org/10.1186/2193-1801-2-393

    Article  PubMed  PubMed Central  Google Scholar 

  64. Kubo T, Ohno Y, Nishino M, Lin P-J, Gautam S, Kauczor H-U et al (2016) Low dose chest CT protocol (50 mAs) as a routine protocol for comprehensive assessment of intrathoracic abnormality. Eur J Radiol Open 3:86–94. https://doi.org/10.1016/j.ejro.2016.04.001

    Article  PubMed  PubMed Central  Google Scholar 

  65. Vachha B, Brodoefel H, Wilcox C, Hackney DB, Moonis G (2013) Radiation dose reduction in soft tissue neck CT using adaptive statistical iterative reconstruction (ASIR). Eur J Radiol 82:2222–2226. https://doi.org/10.1016/j.ejrad.2013.08.014

    Article  PubMed  Google Scholar 

  66. Dangis A, Gieraerts C, Bruecker YD, Janssen L, Valgaeren H, Obbels D, Gillis M, Ranst MV, Frans J, Demeyere AJRCI (2020) Accuracy and reproducibility of low-dose submillisievert chest CT for the diagnosis of COVID-19. Radiology: Cardiothor Imag 2:200196. https://doi.org/10.1148/ryct.2020200196

    Article  Google Scholar 

  67. Kang Z, Li X, Zhou S (2020) Recommendation of low-dose CT in the detection and management of COVID-2019. Eur Radiol 30:4356–4357. https://doi.org/10.1007/s00330-020-06809-6

    Article  CAS  PubMed  Google Scholar 

  68. Tofighi S, Najafi S, Johnston SK, Gholamrezanezhad A (2020) Low-dose CT in COVID-19 outbreak: radiation safety, image wisely, and image gently pledge. Emerg Radiol:1–5. https://doi.org/10.1007/s10140-020-01784-3.

  69. Shiri I, Akhavanallaf A, Sanaat A, Salimi Y, Askari D, Mansouri Z et al (2020) Ultra-low-dose chest CT imaging of COVID-19 patients using a deep residual neural network. Eur Radiol:1–12. https://doi.org/10.1007/s00330-020-07225-6.

Download references

Acknowledgements

None.

Funding

None.

Author information

Authors and Affiliations

Authors

Contributions

J. A. and D. Kh. wrote the main manuscript. Z. S. L., F. E., A. A., and H. B. performed the literature review. B.F. and M. N. designed and edited the main manuscript. The authors read and approved the final manuscript.

Corresponding authors

Correspondence to Bagher Farhood or Masoud Najafi.

Ethics declarations

Ethics approval and consent to participate

Not applicable

Consent for publication

Not applicable

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Azadbakht, J., Khoramian, D., Lajevardi, Z.S. et al. A review on chest CT scanning parameters implemented in COVID-19 patients: bringing low-dose CT protocols into play. Egypt J Radiol Nucl Med 52, 13 (2021). https://doi.org/10.1186/s43055-020-00400-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s43055-020-00400-1

Keywords